This system monitors critical facilities sensors and specialty gas usage on RUMS2.
  • Sensor Status      
  • View all the Graphs      
  • Sensor Configuration & Documentation
  • Return to the NanoLab home page
  • RUMS2 is on-line . Last sampling: 2/23/2012 3:06:59 AM
    Sensor NameDescriptionLow AlarmHigh AlarmCurrent ValueUnitsStatus
    EF_12Lab exhaust alarm (zero indicates no alarm, i.e. exhaust is on). -0.50.50on/off
    EF_34Silane exhaust alarm (zero indicates no alarm, i.e. exhaust is on). -0.50.50on/off
    Low_Level_Gas_WarnLow level gas warning from any zone (zero indicates no warning-level alarms). -0.50.50on/off
    AWN_Chemical_Low_AlarmCheck NaOH and H2SO4 levels on alarm (zero indicates no alarm). -0.50.50on/off
    AWN_General_AlarmState of the Acid Waste Neutralizer general alarm (zero indicates no alarm). -0.50.50on/off
    AWN_Surge_Tank_LevelAcid Waste Neutralizer surge tank level, stage 0. 150450211.6gal
    Compressed_Air_Dew_PointPressure dew point measured at the compressed air source. -112-40-76.82F
    Compressed_Air_General_AlarmGenaral alarm state of the dryer controller (zero indicates no alarm). -0.50.50on/off
    Compressed_Air_N2_BackupState of the CDA N2 backup valve (zero indicates no alarm, i.e. N2 backup is not in use). -0.50.50on/off
    Compressed_Air_PressureThe pressure of the compressed air on the compressor side of the check valve. 90130113.58psi
    DI_General_AlarmState of the DI general alarm (zero indicates no alarm). -0.50.50on/off
    DI_Makeup_FlowRate of flow from the RO system to the DI system. -12517.15gpm
    DI_ResistivityThe final resisitivity of the DI source. 16.518.518.02MOhm-cm
    DI_Tank_LevelDeionized Water tanks level (sum of both tanks). 150023501997.09gal
    PCW_FlowProcess Cooling Water supply flow rate. 35400276.07gpm
    PCW_General_AlarmState of the Process Cooling Water general alarm (zero indicates no alarm). -0.50.50on/off
    PCW_ICW_CrossoverState of the FCV (0 indicates PCW recirc, not crossed over) -0.50.50on/off
    PCW_PressureProcess Cooling Water supply pressure. 5011087.35psi
    PCW_ResistivityResistivity of the PCW loop. 170300243.36kOhm-cm
    PCW_TemperatureProcess Cooling Water supply Temperature. 557061.59F
    Process_VacuumProcess vacuum pressure. 243227.08inHg
    Argon_PressurePre-process argon pressure in 395; pressures less than 40 psi indicate lag bottle is sourcing. 306545.76psi
    N2_Flow_RateFlow rate of N2 at shipping and receiving. 50026001889.19slpm
    N2_PressureN2 pressure at shipping and receiving. 7510091.32psi
    O2_Flow_rateO2 flow rate at shipping and receiving. -5352.1slpm
    O2_PressureO2 pressure at shipping and receiving. 224533.98psi
    Gas_Cabinet_Alarm_L3State of gas cabinets in the L3 gas vault (zero indicates no alarm). -0.50.50on/off
    Gas_Cabinet_Alarm_L5State of gas cabinets in the L5 gas vault (zero indicates no alarm). -0.50.50on/off
    GCAWS_AC_Secondary_H2OWater detection switch in the GCAWS AC trays. -0.50.50on/off
    Lab_3_Relative_HumidityRelative humidity in 384. 207252.53percent
    Lab_3_TemperatureTemperature in 384. 607566.64F
    Lab_5_Relative_HumidityRelative humidity in 584. 207248.19percent
    Lab_5_TemperatureTemperature in 584. 557567.06F
    NH3_599_Weight_lbsWeight of NH3 Cylinder in Rm 599 125180127.75 
    PCW_Pressure_L55th Floor Lab PCW Pressure 408061.62psi
    TestRUMS test channel (for internal use only). -0.50.52.63test